CVE-2026-0611 (GCVE-0-2026-0611)
Vulnerability from cvelistv5 – Published: 2026-06-02 15:39 – Updated: 2026-06-02 19:13
VLAI
Title
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Summary
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
Severity
9.8 (Critical)
SSVC
Exploitation: poc
Automatable: yes
Technical Impact: total
CISA Coordinator (v2.0.3)
CWE
- CWE-306 - Missing Authentication for Critical Function
Assigner
References
3 references
| URL | Tags |
|---|---|
| https://spacelabshealthcare.com/wp-content/upload… | vendor-advisorypatch |
| https://spacelabshealthcare.com/products/diagnost… | product |
| https://www.vulncheck.com/advisories/spacelabs-he… | third-party-advisory |
Impacted products
1 product
| Vendor | Product | Version | |
|---|---|---|---|
| Spacelabs Healthcare | Sentinel |
Affected:
10.5.0 , < 11.6.0
(semver)
|
Date Public
2026-06-02 00:00
Credits
{
"containers": {
"adp": [
{
"metrics": [
{
"other": {
"content": {
"id": "CVE-2026-0611",
"options": [
{
"Exploitation": "poc"
},
{
"Automatable": "yes"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"timestamp": "2026-06-02T19:13:38.769068Z",
"version": "2.0.3"
},
"type": "ssvc"
}
}
],
"providerMetadata": {
"dateUpdated": "2026-06-02T19:13:46.749Z",
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP"
},
"references": [
{
"tags": [
"exploit"
],
"url": "https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"
}
],
"title": "CISA ADP Vulnrichment"
}
],
"cna": {
"affected": [
{
"defaultStatus": "affected",
"product": "Sentinel",
"vendor": "Spacelabs Healthcare",
"versions": [
{
"lessThan": "11.6.0",
"status": "affected",
"version": "10.5.0",
"versionType": "semver"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"value": "Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp."
},
{
"lang": "en",
"type": "finder",
"value": "Jan A. Rodriguez, Pentester, GM Sectec, Corp."
},
{
"lang": "en",
"type": "coordinator",
"value": "VulnCheck"
}
],
"datePublic": "2026-06-02T00:00:00.000Z",
"descriptions": [
{
"lang": "en",
"value": "Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes."
}
],
"metrics": [
{
"cvssV4_0": {
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"attackVector": "NETWORK",
"baseScore": 9.2,
"baseSeverity": "CRITICAL",
"exploitMaturity": "NOT_DEFINED",
"privilegesRequired": "NONE",
"providerUrgency": "NOT_DEFINED",
"subAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"userInteraction": "NONE",
"valueDensity": "NOT_DEFINED",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
"version": "4.0",
"vulnAvailabilityImpact": "HIGH",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "HIGH",
"vulnerabilityResponseEffort": "NOT_DEFINED"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
},
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "NETWORK",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-306",
"description": "Missing Authentication for Critical Function",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"dateUpdated": "2026-06-02T15:39:21.054Z",
"orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"shortName": "VulnCheck"
},
"references": [
{
"tags": [
"vendor-advisory",
"patch"
],
"url": "https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"
},
{
"tags": [
"product"
],
"url": "https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel"
},
{
"tags": [
"third-party-advisory"
],
"url": "https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting"
}
],
"source": {
"discovery": "EXTERNAL"
},
"title": "Spacelabs Healthcare Sentinel 10.5.x \u003c 11.6.0 Unauthenticated RCE via .NET Remoting",
"x_generator": {
"engine": "vulncheck"
}
}
},
"cveMetadata": {
"assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
"assignerShortName": "VulnCheck",
"cveId": "CVE-2026-0611",
"datePublished": "2026-06-02T15:39:21.054Z",
"dateReserved": "2026-01-05T16:55:12.556Z",
"dateUpdated": "2026-06-02T19:13:46.749Z",
"state": "PUBLISHED"
},
"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"vulnerability-lookup:meta": {
"epss": {
"cve": "CVE-2026-0611",
"date": "2026-06-03",
"epss": "0.00189",
"percentile": "0.4054"
},
"nvd": "{\"cve\":{\"id\":\"CVE-2026-0611\",\"sourceIdentifier\":\"disclosure@vulncheck.com\",\"published\":\"2026-06-02T17:16:25.483\",\"lastModified\":\"2026-06-02T20:16:31.633\",\"vulnStatus\":\"Deferred\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.\"}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":9.2,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"PRESENT\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"vulnConfidentialityImpact\":\"HIGH\",\"vulnIntegrityImpact\":\"HIGH\",\"vulnAvailabilityImpact\":\"HIGH\",\"subConfidentialityImpact\":\"NONE\",\"subIntegrityImpact\":\"NONE\",\"subAvailabilityImpact\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirement\":\"NOT_DEFINED\",\"integrityRequirement\":\"NOT_DEFINED\",\"availabilityRequirement\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedVulnIntegrityImpact\":\"NOT_DEFINED\",\"modifiedVulnAvailabilityImpact\":\"NOT_DEFINED\",\"modifiedSubConfidentialityImpact\":\"NOT_DEFINED\",\"modifiedSubIntegrityImpact\":\"NOT_DEFINED\",\"modifiedSubAvailabilityImpact\":\"NOT_DEFINED\",\"Safety\":\"NOT_DEFINED\",\"Automatable\":\"NOT_DEFINED\",\"Recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}],\"cvssMetricV31\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\",\"baseScore\":9.8,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":3.9,\"impactScore\":5.9}]},\"weaknesses\":[{\"source\":\"disclosure@vulncheck.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-306\"}]}],\"references\":[{\"url\":\"https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel\",\"source\":\"disclosure@vulncheck.com\"},{\"url\":\"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf\",\"source\":\"disclosure@vulncheck.com\"},{\"url\":\"https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting\",\"source\":\"disclosure@vulncheck.com\"},{\"url\":\"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf\",\"source\":\"134c704f-9b21-4f2e-91b3-4a467353bcc0\"}]}}",
"vulnrichment": {
"containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2026-0611\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"poc\"}, {\"Automatable\": \"yes\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2026-06-02T19:13:38.769068Z\"}}}], \"references\": [{\"url\": \"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf\", \"tags\": [\"exploit\"]}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2026-06-02T19:01:36.464Z\"}}], \"cna\": {\"title\": \"Spacelabs Healthcare Sentinel 10.5.x \u003c 11.6.0 Unauthenticated RCE via .NET Remoting\", \"source\": {\"discovery\": \"EXTERNAL\"}, \"credits\": [{\"lang\": \"en\", \"type\": \"finder\", \"value\": \"Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp.\"}, {\"lang\": \"en\", \"type\": \"finder\", \"value\": \"Jan A. Rodriguez, Pentester, GM Sectec, Corp.\"}, {\"lang\": \"en\", \"type\": \"coordinator\", \"value\": \"VulnCheck\"}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 9.2, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"NETWORK\", \"baseSeverity\": \"CRITICAL\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N\", \"exploitMaturity\": \"NOT_DEFINED\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"PRESENT\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"HIGH\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"HIGH\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"HIGH\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}, {\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 9.8, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"CRITICAL\", \"vectorString\": \"CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Spacelabs Healthcare\", \"product\": \"Sentinel\", \"versions\": [{\"status\": \"affected\", \"version\": \"10.5.0\", \"lessThan\": \"11.6.0\", \"versionType\": \"semver\"}], \"defaultStatus\": \"affected\"}], \"datePublic\": \"2026-06-02T00:00:00.000Z\", \"references\": [{\"url\": \"https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf\", \"tags\": [\"vendor-advisory\", \"patch\"]}, {\"url\": \"https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel\", \"tags\": [\"product\"]}, {\"url\": \"https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting\", \"tags\": [\"third-party-advisory\"]}], \"x_generator\": {\"engine\": \"vulncheck\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.\"}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-306\", \"description\": \"Missing Authentication for Critical Function\"}]}], \"providerMetadata\": {\"orgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"shortName\": \"VulnCheck\", \"dateUpdated\": \"2026-06-02T15:39:21.054Z\"}}}",
"cveMetadata": "{\"cveId\": \"CVE-2026-0611\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2026-06-02T19:13:46.749Z\", \"dateReserved\": \"2026-01-05T16:55:12.556Z\", \"assignerOrgId\": \"83251b91-4cc7-4094-a5c7-464a1b83ea10\", \"datePublished\": \"2026-06-02T15:39:21.054Z\", \"assignerShortName\": \"VulnCheck\"}",
"dataType": "CVE_RECORD",
"dataVersion": "5.2"
}
}
}
Loading…
Loading…
Experimental. This forecast is provided for visualization only and may change without notice. Do not use it for operational decisions.
Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.
Sightings
| Author | Source | Type | Date | Other |
|---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or observed by the user.
- Confirmed: The vulnerability has been validated from an analyst's perspective.
- Published Proof of Concept: A public proof of concept is available for this vulnerability.
- Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
- Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
- Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
- Not confirmed: The user expressed doubt about the validity of the vulnerability.
- Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.
Loading…
Loading…