CVE-2026-0611 (GCVE-0-2026-0611)

Vulnerability from cvelistv5 – Published: 2026-06-02 15:39 – Updated: 2026-06-02 19:13
VLAI
Title
Spacelabs Healthcare Sentinel 10.5.x < 11.6.0 Unauthenticated RCE via .NET Remoting
Summary
Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.
SSVC
Exploitation: poc Automatable: yes Technical Impact: total
CISA Coordinator (v2.0.3)
CWE
  • CWE-306 - Missing Authentication for Critical Function
Assigner
Impacted products
Vendor Product Version
Spacelabs Healthcare Sentinel Affected: 10.5.0 , < 11.6.0 (semver)
Create a notification for this product.
Date Public
2026-06-02 00:00
Credits
Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp. Jan A. Rodriguez, Pentester, GM Sectec, Corp. VulnCheck
Show details on NVD website

{
  "containers": {
    "adp": [
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2026-0611",
                "options": [
                  {
                    "Exploitation": "poc"
                  },
                  {
                    "Automatable": "yes"
                  },
                  {
                    "Technical Impact": "total"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2026-06-02T19:13:38.769068Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2026-06-02T19:13:46.749Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "references": [
          {
            "tags": [
              "exploit"
            ],
            "url": "https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"
          }
        ],
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "affected",
          "product": "Sentinel",
          "vendor": "Spacelabs Healthcare",
          "versions": [
            {
              "lessThan": "11.6.0",
              "status": "affected",
              "version": "10.5.0",
              "versionType": "semver"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "Victor A. Morales, Senior Pentester Team Leader, GM Sectec, Corp."
        },
        {
          "lang": "en",
          "type": "finder",
          "value": "Jan A. Rodriguez, Pentester, GM Sectec, Corp."
        },
        {
          "lang": "en",
          "type": "coordinator",
          "value": "VulnCheck"
        }
      ],
      "datePublic": "2026-06-02T00:00:00.000Z",
      "descriptions": [
        {
          "lang": "en",
          "value": "Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "PRESENT",
            "attackVector": "NETWORK",
            "baseScore": 9.2,
            "baseSeverity": "CRITICAL",
            "exploitMaturity": "NOT_DEFINED",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "HIGH",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "HIGH",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "LOW",
            "attackVector": "NETWORK",
            "availabilityImpact": "HIGH",
            "baseScore": 9.8,
            "baseSeverity": "CRITICAL",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "HIGH",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-306",
              "description": "Missing Authentication for Critical Function",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2026-06-02T15:39:21.054Z",
        "orgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
        "shortName": "VulnCheck"
      },
      "references": [
        {
          "tags": [
            "vendor-advisory",
            "patch"
          ],
          "url": "https://spacelabshealthcare.com/wp-content/uploads/2026/06/079-0273-00-RevA-Security-Advisory-Sentinel-.NET-Remoting-Vulnerability.pdf"
        },
        {
          "tags": [
            "product"
          ],
          "url": "https://spacelabshealthcare.com/products/diagnostic-cardiology/connectivity/sentinel"
        },
        {
          "tags": [
            "third-party-advisory"
          ],
          "url": "https://www.vulncheck.com/advisories/spacelabs-healthcare-sentinel-x-unauthenticated-rce-via-net-remoting"
        }
      ],
      "source": {
        "discovery": "EXTERNAL"
      },
      "title": "Spacelabs Healthcare Sentinel 10.5.x \u003c 11.6.0 Unauthenticated RCE via .NET Remoting",
      "x_generator": {
        "engine": "vulncheck"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "83251b91-4cc7-4094-a5c7-464a1b83ea10",
    "assignerShortName": "VulnCheck",
    "cveId": "CVE-2026-0611",
    "datePublished": "2026-06-02T15:39:21.054Z",
    "dateReserved": "2026-01-05T16:55:12.556Z",
    "dateUpdated": "2026-06-02T19:13:46.749Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.2"
}