fkie_cve-2024-48974
Vulnerability from fkie_nvd
Published
2024-11-14 22:15
Modified
2024-11-15 13:58
Severity ?
Summary
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Impacted products
Vendor Product Version



{
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device\u0027s configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure."
    },
    {
      "lang": "es",
      "value": "El respirador no realiza comprobaciones adecuadas de la integridad de los archivos al adoptar actualizaciones de firmware. Esto hace posible que un atacante fuerce cambios no autorizados en la configuraci\u00f3n del dispositivo o comprometa la funcionalidad del dispositivo al introducir un archivo de firmware comprometido o ileg\u00edtimo. Esto podr\u00eda interrumpir el funcionamiento del dispositivo o provocar la divulgaci\u00f3n no autorizada de informaci\u00f3n."
    }
  ],
  "id": "CVE-2024-48974",
  "lastModified": "2024-11-15T13:58:08.913",
  "metrics": {
    "cvssMetricV31": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "LOCAL",
          "availabilityImpact": "HIGH",
          "baseScore": 9.3,
          "baseSeverity": "CRITICAL",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "HIGH",
          "privilegesRequired": "NONE",
          "scope": "CHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
          "version": "3.1"
        },
        "exploitabilityScore": 2.5,
        "impactScore": 6.0,
        "source": "productsecurity@baxter.com",
        "type": "Secondary"
      }
    ]
  },
  "published": "2024-11-14T22:15:18.727",
  "references": [
    {
      "source": "productsecurity@baxter.com",
      "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
    }
  ],
  "sourceIdentifier": "productsecurity@baxter.com",
  "vulnStatus": "Awaiting Analysis",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-494"
        }
      ],
      "source": "productsecurity@baxter.com",
      "type": "Secondary"
    }
  ]
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.