cve-2024-48974
Vulnerability from cvelistv5
Published
2024-11-14 21:27
Modified
2024-11-18 15:32
Severity ?
Summary
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
Impacted products
Vendor Product Version
Baxter Life2000 Ventilation System Version: 06.08.00.00 and prior
Create a notification for this product.
Show details on NVD website


{
   containers: {
      adp: [
         {
            affected: [
               {
                  cpes: [
                     "cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*",
                  ],
                  defaultStatus: "unaffected",
                  product: "life2000_ventilator_firmware",
                  vendor: "baxter",
                  versions: [
                     {
                        lessThanOrEqual: "06.08.00.00",
                        status: "affected",
                        version: "0",
                        versionType: "custom",
                     },
                  ],
               },
            ],
            metrics: [
               {
                  other: {
                     content: {
                        id: "CVE-2024-48974",
                        options: [
                           {
                              Exploitation: "none",
                           },
                           {
                              Automatable: "no",
                           },
                           {
                              "Technical Impact": "total",
                           },
                        ],
                        role: "CISA Coordinator",
                        timestamp: "2024-11-18T15:31:47.610421Z",
                        version: "2.0.3",
                     },
                     type: "ssvc",
                  },
               },
            ],
            providerMetadata: {
               dateUpdated: "2024-11-18T15:32:38.488Z",
               orgId: "134c704f-9b21-4f2e-91b3-4a467353bcc0",
               shortName: "CISA-ADP",
            },
            title: "CISA ADP Vulnrichment",
         },
      ],
      cna: {
         affected: [
            {
               defaultStatus: "unaffected",
               product: "Life2000 Ventilation System",
               vendor: "Baxter",
               versions: [
                  {
                     status: "affected",
                     version: "06.08.00.00 and prior",
                  },
               ],
            },
         ],
         datePublic: "2024-11-14T21:50:00.000Z",
         descriptions: [
            {
               lang: "en",
               supportingMedia: [
                  {
                     base64: false,
                     type: "text/html",
                     value: "The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.",
                  },
               ],
               value: "The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.",
            },
         ],
         impacts: [
            {
               capecId: "CAPEC-186",
               descriptions: [
                  {
                     lang: "en",
                     value: "CAPEC-186 Malicious Software Update",
                  },
               ],
            },
            {
               capecId: "CAPEC-117",
               descriptions: [
                  {
                     lang: "en",
                     value: "CAPEC-117 Interception",
                  },
               ],
            },
         ],
         metrics: [
            {
               cvssV3_1: {
                  attackComplexity: "LOW",
                  attackVector: "LOCAL",
                  availabilityImpact: "HIGH",
                  baseScore: 9.3,
                  baseSeverity: "CRITICAL",
                  confidentialityImpact: "HIGH",
                  integrityImpact: "HIGH",
                  privilegesRequired: "NONE",
                  scope: "CHANGED",
                  userInteraction: "NONE",
                  vectorString: "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
                  version: "3.1",
               },
               format: "CVSS",
               scenarios: [
                  {
                     lang: "en",
                     value: "GENERAL",
                  },
               ],
            },
         ],
         problemTypes: [
            {
               descriptions: [
                  {
                     cweId: "CWE-494",
                     description: "CWE-494 Download of Code Without Integrity Check",
                     lang: "en",
                     type: "CWE",
                  },
               ],
            },
         ],
         providerMetadata: {
            dateUpdated: "2024-11-14T21:50:36.915Z",
            orgId: "dba971b9-eb30-4121-91e1-3b45611354aa",
            shortName: "Baxter",
         },
         references: [
            {
               url: "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
            },
         ],
         source: {
            discovery: "UNKNOWN",
         },
         title: "Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates",
         x_generator: {
            engine: "Vulnogram 0.2.0",
         },
      },
   },
   cveMetadata: {
      assignerOrgId: "dba971b9-eb30-4121-91e1-3b45611354aa",
      assignerShortName: "Baxter",
      cveId: "CVE-2024-48974",
      datePublished: "2024-11-14T21:27:10.284Z",
      dateReserved: "2024-10-10T19:24:41.495Z",
      dateUpdated: "2024-11-18T15:32:38.488Z",
      state: "PUBLISHED",
   },
   dataType: "CVE_RECORD",
   dataVersion: "5.1",
   "vulnerability-lookup:meta": {
      nvd: "{\"cve\":{\"id\":\"CVE-2024-48974\",\"sourceIdentifier\":\"productsecurity@baxter.com\",\"published\":\"2024-11-14T22:15:18.727\",\"lastModified\":\"2024-11-15T13:58:08.913\",\"vulnStatus\":\"Awaiting Analysis\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.\"},{\"lang\":\"es\",\"value\":\"El respirador no realiza comprobaciones adecuadas de la integridad de los archivos al adoptar actualizaciones de firmware. Esto hace posible que un atacante fuerce cambios no autorizados en la configuración del dispositivo o comprometa la funcionalidad del dispositivo al introducir un archivo de firmware comprometido o ilegítimo. Esto podría interrumpir el funcionamiento del dispositivo o provocar la divulgación no autorizada de información.\"}],\"metrics\":{\"cvssMetricV31\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\",\"baseScore\":9.3,\"baseSeverity\":\"CRITICAL\",\"attackVector\":\"LOCAL\",\"attackComplexity\":\"LOW\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"CHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"HIGH\",\"availabilityImpact\":\"HIGH\"},\"exploitabilityScore\":2.5,\"impactScore\":6.0}]},\"weaknesses\":[{\"source\":\"productsecurity@baxter.com\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-494\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\",\"source\":\"productsecurity@baxter.com\"}]}}",
      vulnrichment: {
         containers: "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2024-48974\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2024-11-18T15:31:47.610421Z\"}}}], \"affected\": [{\"cpes\": [\"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*\"], \"vendor\": \"baxter\", \"product\": \"life2000_ventilator_firmware\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\", \"versionType\": \"custom\", \"lessThanOrEqual\": \"06.08.00.00\"}], \"defaultStatus\": \"unaffected\"}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2024-11-18T15:32:27.939Z\"}}], \"cna\": {\"title\": \"Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates\", \"source\": {\"discovery\": \"UNKNOWN\"}, \"impacts\": [{\"capecId\": \"CAPEC-186\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-186 Malicious Software Update\"}]}, {\"capecId\": \"CAPEC-117\", \"descriptions\": [{\"lang\": \"en\", \"value\": \"CAPEC-117 Interception\"}]}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"CHANGED\", \"version\": \"3.1\", \"baseScore\": 9.3, \"attackVector\": \"LOCAL\", \"baseSeverity\": \"CRITICAL\", \"vectorString\": \"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H\", \"integrityImpact\": \"HIGH\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"availabilityImpact\": \"HIGH\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Baxter\", \"product\": \"Life2000 Ventilation System\", \"versions\": [{\"status\": \"affected\", \"version\": \"06.08.00.00 and prior\"}], \"defaultStatus\": \"unaffected\"}], \"datePublic\": \"2024-11-14T21:50:00.000Z\", \"references\": [{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-494\", \"description\": \"CWE-494 Download of Code Without Integrity Check\"}]}], \"providerMetadata\": {\"orgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"shortName\": \"Baxter\", \"dateUpdated\": \"2024-11-14T21:50:36.915Z\"}}}",
         cveMetadata: "{\"cveId\": \"CVE-2024-48974\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2024-11-18T15:32:38.488Z\", \"dateReserved\": \"2024-10-10T19:24:41.495Z\", \"assignerOrgId\": \"dba971b9-eb30-4121-91e1-3b45611354aa\", \"datePublished\": \"2024-11-14T21:27:10.284Z\", \"assignerShortName\": \"Baxter\"}",
         dataType: "CVE_RECORD",
         dataVersion: "5.1",
      },
   },
}


Log in or create an account to share your comment.

Security Advisory comment format.

This schema specifies the format of a comment related to a security advisory.

UUIDv4 of the comment
UUIDv4 of the Vulnerability-Lookup instance
When the comment was created originally
When the comment was last updated
Title of the comment
Description of the comment
The identifier of the vulnerability (CVE ID, GHSA-ID, PYSEC ID, etc.).



Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.