cve-2025-0683
Vulnerability from cvelistv5
Published
2025-01-30 18:17
Modified
2025-02-12 20:41
Summary
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Impacted products
Show details on NVD website


{
  "containers": {
    "adp": [
      {
        "providerMetadata": {
          "dateUpdated": "2025-01-31T15:35:10.389Z",
          "orgId": "af854a3a-2127-422b-91ae-364da2661108",
          "shortName": "CVE"
        },
        "references": [
          {
            "url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor"
          },
          {
            "url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
          },
          {
            "url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/"
          }
        ],
        "title": "CVE Program Container",
        "x_generator": {
          "engine": "ADPogram 0.0.1"
        }
      },
      {
        "metrics": [
          {
            "other": {
              "content": {
                "id": "CVE-2025-0683",
                "options": [
                  {
                    "Exploitation": "none"
                  },
                  {
                    "Automatable": "no"
                  },
                  {
                    "Technical Impact": "partial"
                  }
                ],
                "role": "CISA Coordinator",
                "timestamp": "2025-01-30T19:06:59.525559Z",
                "version": "2.0.3"
              },
              "type": "ssvc"
            }
          }
        ],
        "providerMetadata": {
          "dateUpdated": "2025-02-12T20:41:36.789Z",
          "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
          "shortName": "CISA-ADP"
        },
        "title": "CISA ADP Vulnrichment"
      }
    ],
    "cna": {
      "affected": [
        {
          "defaultStatus": "unaffected",
          "product": "CMS8000 Patient Monitor",
          "vendor": "Contec Health",
          "versions": [
            {
              "status": "affected",
              "version": "All versions"
            }
          ]
        }
      ],
      "credits": [
        {
          "lang": "en",
          "type": "finder",
          "value": "An anonymous researcher reported these vulnerabilities to CISA."
        }
      ],
      "descriptions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
            }
          ],
          "value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
        }
      ],
      "metrics": [
        {
          "cvssV4_0": {
            "Automatable": "NOT_DEFINED",
            "Recovery": "NOT_DEFINED",
            "Safety": "NOT_DEFINED",
            "attackComplexity": "LOW",
            "attackRequirements": "PRESENT",
            "attackVector": "NETWORK",
            "baseScore": 8.2,
            "baseSeverity": "HIGH",
            "privilegesRequired": "NONE",
            "providerUrgency": "NOT_DEFINED",
            "subAvailabilityImpact": "NONE",
            "subConfidentialityImpact": "NONE",
            "subIntegrityImpact": "NONE",
            "userInteraction": "NONE",
            "valueDensity": "NOT_DEFINED",
            "vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N",
            "version": "4.0",
            "vulnAvailabilityImpact": "NONE",
            "vulnConfidentialityImpact": "HIGH",
            "vulnIntegrityImpact": "NONE",
            "vulnerabilityResponseEffort": "NOT_DEFINED"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        },
        {
          "cvssV3_1": {
            "attackComplexity": "HIGH",
            "attackVector": "NETWORK",
            "availabilityImpact": "NONE",
            "baseScore": 5.9,
            "baseSeverity": "MEDIUM",
            "confidentialityImpact": "HIGH",
            "integrityImpact": "NONE",
            "privilegesRequired": "NONE",
            "scope": "UNCHANGED",
            "userInteraction": "NONE",
            "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
            "version": "3.1"
          },
          "format": "CVSS",
          "scenarios": [
            {
              "lang": "en",
              "value": "GENERAL"
            }
          ]
        }
      ],
      "problemTypes": [
        {
          "descriptions": [
            {
              "cweId": "CWE-359",
              "description": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
              "lang": "en",
              "type": "CWE"
            }
          ]
        }
      ],
      "providerMetadata": {
        "dateUpdated": "2025-01-31T17:07:47.786Z",
        "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
        "shortName": "icscert"
      },
      "references": [
        {
          "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
        },
        {
          "url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
        }
      ],
      "solutions": [
        {
          "lang": "en",
          "supportingMedia": [
            {
              "base64": false,
              "type": "text/html",
              "value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\n\n\u003cbr\u003e"
            }
          ],
          "value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s  safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
        }
      ],
      "source": {
        "advisory": "ICSMA-25-030-01",
        "discovery": "EXTERNAL"
      },
      "title": "Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor",
      "x_generator": {
        "engine": "Vulnogram 0.2.0"
      }
    }
  },
  "cveMetadata": {
    "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
    "assignerShortName": "icscert",
    "cveId": "CVE-2025-0683",
    "datePublished": "2025-01-30T18:17:35.867Z",
    "dateReserved": "2025-01-23T18:11:20.770Z",
    "dateUpdated": "2025-02-12T20:41:36.789Z",
    "state": "PUBLISHED"
  },
  "dataType": "CVE_RECORD",
  "dataVersion": "5.1",
  "vulnerability-lookup:meta": {
    "nvd": "{\"cve\":{\"id\":\"CVE-2025-0683\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2025-01-30T19:15:14.460\",\"lastModified\":\"2025-01-31T18:15:37.750\",\"vulnStatus\":\"Received\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \\npatient data to a hard-coded public IP address when a patient is hooked \\nup to the monitor. This could lead to a leakage of confidential patient \\ndata to any device with that IP address or an attacker in a \\nmachine-in-the-middle scenario.\"},{\"lang\":\"es\",\"value\":\"En su configuraci\u00f3n predeterminada, el producto afectado transmite datos del paciente en texto plano a una direcci\u00f3n IP p\u00fablica codificada cuando un paciente est\u00e1 conectado al monitor. Esto podr\u00eda provocar una fuga de datos confidenciales del paciente a cualquier dispositivo con esa direcci\u00f3n IP o a un atacante en un escenario de m\u00e1quina intermedia.\"}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":8.2,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"PRESENT\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"vulnerableSystemConfidentiality\":\"HIGH\",\"vulnerableSystemIntegrity\":\"NONE\",\"vulnerableSystemAvailability\":\"NONE\",\"subsequentSystemConfidentiality\":\"NONE\",\"subsequentSystemIntegrity\":\"NONE\",\"subsequentSystemAvailability\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirements\":\"NOT_DEFINED\",\"integrityRequirements\":\"NOT_DEFINED\",\"availabilityRequirements\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnerableSystemConfidentiality\":\"NOT_DEFINED\",\"modifiedVulnerableSystemIntegrity\":\"NOT_DEFINED\",\"modifiedVulnerableSystemAvailability\":\"NOT_DEFINED\",\"modifiedSubsequentSystemConfidentiality\":\"NOT_DEFINED\",\"modifiedSubsequentSystemIntegrity\":\"NOT_DEFINED\",\"modifiedSubsequentSystemAvailability\":\"NOT_DEFINED\",\"safety\":\"NOT_DEFINED\",\"automatable\":\"NOT_DEFINED\",\"recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}],\"cvssMetricV31\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"3.1\",\"vectorString\":\"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N\",\"baseScore\":5.9,\"baseSeverity\":\"MEDIUM\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"HIGH\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"NONE\",\"scope\":\"UNCHANGED\",\"confidentialityImpact\":\"HIGH\",\"integrityImpact\":\"NONE\",\"availabilityImpact\":\"NONE\"},\"exploitabilityScore\":2.2,\"impactScore\":3.6}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-359\"}]}],\"references\":[{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\"},{\"url\":\"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\"},{\"url\":\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\",\"source\":\"af854a3a-2127-422b-91ae-364da2661108\"}]}}",
    "vulnrichment": {
      "containers": "{\"adp\": [{\"title\": \"CVE Program Container\", \"references\": [{\"url\": \"https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor\"}, {\"url\": \"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"}, {\"url\": \"https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/\"}], \"x_generator\": {\"engine\": \"ADPogram 0.0.1\"}, \"providerMetadata\": {\"orgId\": \"af854a3a-2127-422b-91ae-364da2661108\", \"shortName\": \"CVE\", \"dateUpdated\": \"2025-01-31T15:35:10.389Z\"}}, {\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2025-0683\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"none\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"partial\"}], \"version\": \"2.0.3\", \"timestamp\": \"2025-01-30T19:06:59.525559Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2025-02-12T20:40:17.373Z\"}}], \"cna\": {\"title\": \"Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor\", \"source\": {\"advisory\": \"ICSMA-25-030-01\", \"discovery\": \"EXTERNAL\"}, \"credits\": [{\"lang\": \"en\", \"type\": \"finder\", \"value\": \"An anonymous researcher reported these vulnerabilities to CISA.\"}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 8.2, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"PRESENT\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"NONE\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"NONE\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"HIGH\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}, {\"format\": \"CVSS\", \"cvssV3_1\": {\"scope\": \"UNCHANGED\", \"version\": \"3.1\", \"baseScore\": 5.9, \"attackVector\": \"NETWORK\", \"baseSeverity\": \"MEDIUM\", \"vectorString\": \"CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N\", \"integrityImpact\": \"NONE\", \"userInteraction\": \"NONE\", \"attackComplexity\": \"HIGH\", \"availabilityImpact\": \"NONE\", \"privilegesRequired\": \"NONE\", \"confidentialityImpact\": \"HIGH\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Contec Health\", \"product\": \"CMS8000 Patient Monitor\", \"versions\": [{\"status\": \"affected\", \"version\": \"All versions\"}], \"defaultStatus\": \"unaffected\"}], \"solutions\": [{\"lang\": \"en\", \"value\": \"Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\\n\\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s  safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication .\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003ePlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA\u0027s \u003ca target=\\\"_blank\\\" rel=\\\"nofollow\\\" href=\\\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\\\"\u003esafety communication\u003c/a\u003e.\u003c/p\u003e\\n\\n\u003cbr\u003e\", \"base64\": false}]}], \"references\": [{\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01\"}, {\"url\": \"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\"}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \\npatient data to a hard-coded public IP address when a patient is hooked \\nup to the monitor. This could lead to a leakage of confidential patient \\ndata to any device with that IP address or an attacker in a \\nmachine-in-the-middle scenario.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \\npatient data to a hard-coded public IP address when a patient is hooked \\nup to the monitor. This could lead to a leakage of confidential patient \\ndata to any device with that IP address or an attacker in a \\nmachine-in-the-middle scenario.\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-359\", \"description\": \"CWE-359 Exposure of Private Personal Information to an Unauthorized Actor\"}]}], \"providerMetadata\": {\"orgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"shortName\": \"icscert\", \"dateUpdated\": \"2025-01-31T17:07:47.786Z\"}}}",
      "cveMetadata": "{\"cveId\": \"CVE-2025-0683\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2025-02-12T20:41:36.789Z\", \"dateReserved\": \"2025-01-23T18:11:20.770Z\", \"assignerOrgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"datePublished\": \"2025-01-30T18:17:35.867Z\", \"assignerShortName\": \"icscert\"}",
      "dataType": "CVE_RECORD",
      "dataVersion": "5.1"
    }
  }
}


Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Sightings

Author Source Type Date

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
  • Confirmed: The vulnerability is confirmed from an analyst perspective.
  • Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
  • Patched: This vulnerability was successfully patched by the user reporting the sighting.
  • Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
  • Not confirmed: The user expresses doubt about the veracity of the vulnerability.
  • Not patched: This vulnerability was not successfully patched by the user reporting the sighting.