CVE-2025-1204
Vulnerability from cvelistv5
Published
2025-02-25 17:00
Modified
2025-02-25 17:23
Severity ?
EPSS score ?
Summary
The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.
References
Impacted products
Vendor | Product | Version | ||
---|---|---|---|---|
Contec Health | CMS8000 Patient Monitor |
Version: 0 |
{ "containers": { "adp": [ { "metrics": [ { "other": { "content": { "id": "CVE-2025-1204", "options": [ { "Exploitation": "poc" }, { "Automatable": "no" }, { "Technical Impact": "total" } ], "role": "CISA Coordinator", "timestamp": "2025-02-25T17:22:58.082071Z", "version": "2.0.3" }, "type": "ssvc" } } ], "providerMetadata": { "dateUpdated": "2025-02-25T17:23:02.222Z", "orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP" }, "title": "CISA ADP Vulnrichment" } ], "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "CMS8000 Patient Monitor", "vendor": "Contec Health", "versions": [ { "status": "affected", "version": "0" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Claroty - Team82" } ], "datePublic": "2025-02-25T17:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device." } ], "value": "The \"update\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device." } ], "metrics": [ { "cvssV4_0": { "Automatable": "NOT_DEFINED", "Recovery": "NOT_DEFINED", "Safety": "NOT_DEFINED", "attackComplexity": "LOW", "attackRequirements": "PRESENT", "attackVector": "NETWORK", "baseScore": 7.7, "baseSeverity": "HIGH", "privilegesRequired": "NONE", "providerUrgency": "NOT_DEFINED", "subAvailabilityImpact": "NONE", "subConfidentialityImpact": "NONE", "subIntegrityImpact": "NONE", "userInteraction": "PASSIVE", "valueDensity": "NOT_DEFINED", "vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N", "version": "4.0", "vulnAvailabilityImpact": "HIGH", "vulnConfidentialityImpact": "HIGH", "vulnIntegrityImpact": "HIGH", "vulnerabilityResponseEffort": "NOT_DEFINED" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-912", "description": "CWE-912 Hidden Functionality", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "dateUpdated": "2025-02-25T17:00:16.133Z", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert" }, "references": [ { "tags": [ "third-party-advisory" ], "url": "https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org" }, { "tags": [ "government-resource" ], "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01" } ], "source": { "discovery": "UNKNOWN" }, "workarounds": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e" } ], "value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120." } ], "x_generator": { "engine": "Vulnogram 0.2.0" } } }, "cveMetadata": { "assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2025-1204", "datePublished": "2025-02-25T17:00:16.133Z", "dateReserved": "2025-02-10T17:00:02.210Z", "dateUpdated": "2025-02-25T17:23:02.222Z", "state": "PUBLISHED" }, "dataType": "CVE_RECORD", "dataVersion": "5.1", "vulnerability-lookup:meta": { "nvd": "{\"cve\":{\"id\":\"CVE-2025-1204\",\"sourceIdentifier\":\"ics-cert@hq.dhs.gov\",\"published\":\"2025-02-25T17:15:14.057\",\"lastModified\":\"2025-02-25T17:15:14.057\",\"vulnStatus\":\"Received\",\"cveTags\":[],\"descriptions\":[{\"lang\":\"en\",\"value\":\"The \\\"update\\\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.\"}],\"metrics\":{\"cvssMetricV40\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Secondary\",\"cvssData\":{\"version\":\"4.0\",\"vectorString\":\"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X\",\"baseScore\":7.7,\"baseSeverity\":\"HIGH\",\"attackVector\":\"NETWORK\",\"attackComplexity\":\"LOW\",\"attackRequirements\":\"PRESENT\",\"privilegesRequired\":\"NONE\",\"userInteraction\":\"PASSIVE\",\"vulnerableSystemConfidentiality\":\"HIGH\",\"vulnerableSystemIntegrity\":\"HIGH\",\"vulnerableSystemAvailability\":\"HIGH\",\"subsequentSystemConfidentiality\":\"NONE\",\"subsequentSystemIntegrity\":\"NONE\",\"subsequentSystemAvailability\":\"NONE\",\"exploitMaturity\":\"NOT_DEFINED\",\"confidentialityRequirements\":\"NOT_DEFINED\",\"integrityRequirements\":\"NOT_DEFINED\",\"availabilityRequirements\":\"NOT_DEFINED\",\"modifiedAttackVector\":\"NOT_DEFINED\",\"modifiedAttackComplexity\":\"NOT_DEFINED\",\"modifiedAttackRequirements\":\"NOT_DEFINED\",\"modifiedPrivilegesRequired\":\"NOT_DEFINED\",\"modifiedUserInteraction\":\"NOT_DEFINED\",\"modifiedVulnerableSystemConfidentiality\":\"NOT_DEFINED\",\"modifiedVulnerableSystemIntegrity\":\"NOT_DEFINED\",\"modifiedVulnerableSystemAvailability\":\"NOT_DEFINED\",\"modifiedSubsequentSystemConfidentiality\":\"NOT_DEFINED\",\"modifiedSubsequentSystemIntegrity\":\"NOT_DEFINED\",\"modifiedSubsequentSystemAvailability\":\"NOT_DEFINED\",\"safety\":\"NOT_DEFINED\",\"automatable\":\"NOT_DEFINED\",\"recovery\":\"NOT_DEFINED\",\"valueDensity\":\"NOT_DEFINED\",\"vulnerabilityResponseEffort\":\"NOT_DEFINED\",\"providerUrgency\":\"NOT_DEFINED\"}}]},\"weaknesses\":[{\"source\":\"ics-cert@hq.dhs.gov\",\"type\":\"Primary\",\"description\":[{\"lang\":\"en\",\"value\":\"CWE-912\"}]}],\"references\":[{\"url\":\"https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org\",\"source\":\"ics-cert@hq.dhs.gov\"},{\"url\":\"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01\",\"source\":\"ics-cert@hq.dhs.gov\"}]}}", "vulnrichment": { "containers": "{\"adp\": [{\"title\": \"CISA ADP Vulnrichment\", \"metrics\": [{\"other\": {\"type\": \"ssvc\", \"content\": {\"id\": \"CVE-2025-1204\", \"role\": \"CISA Coordinator\", \"options\": [{\"Exploitation\": \"poc\"}, {\"Automatable\": \"no\"}, {\"Technical Impact\": \"total\"}], \"version\": \"2.0.3\", \"timestamp\": \"2025-02-25T17:22:58.082071Z\"}}}], \"providerMetadata\": {\"orgId\": \"134c704f-9b21-4f2e-91b3-4a467353bcc0\", \"shortName\": \"CISA-ADP\", \"dateUpdated\": \"2025-02-25T17:22:29.789Z\"}}], \"cna\": {\"source\": {\"discovery\": \"UNKNOWN\"}, \"credits\": [{\"lang\": \"en\", \"type\": \"finder\", \"value\": \"Claroty - Team82\"}], \"metrics\": [{\"format\": \"CVSS\", \"cvssV4_0\": {\"Safety\": \"NOT_DEFINED\", \"version\": \"4.0\", \"Recovery\": \"NOT_DEFINED\", \"baseScore\": 7.7, \"Automatable\": \"NOT_DEFINED\", \"attackVector\": \"NETWORK\", \"baseSeverity\": \"HIGH\", \"valueDensity\": \"NOT_DEFINED\", \"vectorString\": \"CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N\", \"providerUrgency\": \"NOT_DEFINED\", \"userInteraction\": \"PASSIVE\", \"attackComplexity\": \"LOW\", \"attackRequirements\": \"PRESENT\", \"privilegesRequired\": \"NONE\", \"subIntegrityImpact\": \"NONE\", \"vulnIntegrityImpact\": \"HIGH\", \"subAvailabilityImpact\": \"NONE\", \"vulnAvailabilityImpact\": \"HIGH\", \"subConfidentialityImpact\": \"NONE\", \"vulnConfidentialityImpact\": \"HIGH\", \"vulnerabilityResponseEffort\": \"NOT_DEFINED\"}, \"scenarios\": [{\"lang\": \"en\", \"value\": \"GENERAL\"}]}], \"affected\": [{\"vendor\": \"Contec Health\", \"product\": \"CMS8000 Patient Monitor\", \"versions\": [{\"status\": \"affected\", \"version\": \"0\"}], \"defaultStatus\": \"unaffected\"}], \"datePublic\": \"2025-02-25T17:00:00.000Z\", \"references\": [{\"url\": \"https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org\", \"tags\": [\"third-party-advisory\"]}, {\"url\": \"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01\", \"tags\": [\"government-resource\"]}], \"workarounds\": [{\"lang\": \"en\", \"value\": \"Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\\n\\nIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"\u003cp\u003ePer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\u003c/p\u003e\u003cp\u003eIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.\u003c/p\u003e\", \"base64\": false}]}], \"x_generator\": {\"engine\": \"Vulnogram 0.2.0\"}, \"descriptions\": [{\"lang\": \"en\", \"value\": \"The \\\"update\\\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.\", \"supportingMedia\": [{\"type\": \"text/html\", \"value\": \"The \\\"update\\\" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the \u0027C\u0027 button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.\", \"base64\": false}]}], \"problemTypes\": [{\"descriptions\": [{\"lang\": \"en\", \"type\": \"CWE\", \"cweId\": \"CWE-912\", \"description\": \"CWE-912 Hidden Functionality\"}]}], \"providerMetadata\": {\"orgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"shortName\": \"icscert\", \"dateUpdated\": \"2025-02-25T17:00:16.133Z\"}}}", "cveMetadata": "{\"cveId\": \"CVE-2025-1204\", \"state\": \"PUBLISHED\", \"dateUpdated\": \"2025-02-25T17:23:02.222Z\", \"dateReserved\": \"2025-02-10T17:00:02.210Z\", \"assignerOrgId\": \"7d14cffa-0d7d-4270-9dc0-52cabd5a23a6\", \"datePublished\": \"2025-02-25T17:00:16.133Z\", \"assignerShortName\": \"icscert\"}", "dataType": "CVE_RECORD", "dataVersion": "5.1" } } }
Loading…
Loading…
Sightings
Author | Source | Type | Date |
---|
Nomenclature
- Seen: The vulnerability was mentioned, discussed, or seen somewhere by the user.
- Confirmed: The vulnerability is confirmed from an analyst perspective.
- Exploited: This vulnerability was exploited and seen by the user reporting the sighting.
- Patched: This vulnerability was successfully patched by the user reporting the sighting.
- Not exploited: This vulnerability was not exploited or seen by the user reporting the sighting.
- Not confirmed: The user expresses doubt about the veracity of the vulnerability.
- Not patched: This vulnerability was not successfully patched by the user reporting the sighting.