GCVE Workshop - 22 September 2026 (14:00-18:00), Luxembourg Before The Vulnopticon Conference - Registration

FKIE_CVE-2015-3952

Vulnerability from fkie_nvd - Published: 2019-03-25 16:29 - Updated: 2026-06-17 00:26
Summary
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
ics-cert@hq.dhs.govhttps://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource
af854a3a-2127-422b-91ae-364da2661108https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01Mitigation, Third Party Advisory, US Government Resource

{
  "affected": [
    {
      "affectedData": [
        {
          "product": "Plum A+ Infusion System",
          "vendor": "Hospira",
          "versions": [
            {
              "status": "affected",
              "version": "\u003c= 13.4"
            }
          ]
        },
        {
          "product": "Plum A+3 Infusion System",
          "vendor": "Hospira",
          "versions": [
            {
              "status": "affected",
              "version": "\u003c= 13.6"
            }
          ]
        },
        {
          "product": "Symbiq Infusion System",
          "vendor": "Hospira",
          "versions": [
            {
              "status": "affected",
              "version": "\u003c= 3.13"
            }
          ]
        }
      ],
      "source": "ics-cert@hq.dhs.gov"
    }
  ],
  "configurations": [
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2",
              "versionEndIncluding": "13.4",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+_infusion_system:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "1DB2490B-0318-4770-BF45-CD7527F15D7F",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:plum_a\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "8536E705-89E8-47CB-9567-6AD65FBA0F1B",
              "versionEndIncluding": "13.6",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:plum_a\\+3_infusion_system:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "423AA561-8E38-4378-814B-1008B96F27A6",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    },
    {
      "nodes": [
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*",
              "matchCriteriaId": "C6DD5141-72AB-4694-8595-B4BED8EC7773",
              "versionEndIncluding": "3.13",
              "vulnerable": true
            }
          ],
          "negate": false,
          "operator": "OR"
        },
        {
          "cpeMatch": [
            {
              "criteria": "cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*",
              "matchCriteriaId": "5C295E1A-BF60-476D-B972-5C5C28D7633B",
              "vulnerable": false
            }
          ],
          "negate": false,
          "operator": "OR"
        }
      ],
      "operator": "AND"
    }
  ],
  "cveTags": [],
  "descriptions": [
    {
      "lang": "en",
      "value": "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
    },
    {
      "lang": "es",
      "value": "Las claves inal\u00e1mbricas se almacenan como texto en claro en Hospira Plum A+ Infusion System en su versi\u00f3n 13.4 y anteriores, Plum A+3 Infusion System en su versi\u00f3n 13.6 y anteriores y Symbiq Infusion System, en su versi\u00f3n 3.13 y anteriores. Hospira recomienda con los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado tambi\u00e9n el Plum 360 Infusion System que no es vulnerable a este problema."
    }
  ],
  "id": "CVE-2015-3952",
  "lastModified": "2026-06-17T00:26:31.893",
  "metrics": {
    "cvssMetricV2": [
      {
        "acInsufInfo": false,
        "baseSeverity": "MEDIUM",
        "cvssData": {
          "accessComplexity": "LOW",
          "accessVector": "NETWORK",
          "authentication": "NONE",
          "availabilityImpact": "NONE",
          "baseScore": 5.0,
          "confidentialityImpact": "PARTIAL",
          "integrityImpact": "NONE",
          "vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
          "version": "2.0"
        },
        "exploitabilityScore": 10.0,
        "impactScore": 2.9,
        "obtainAllPrivilege": false,
        "obtainOtherPrivilege": false,
        "obtainUserPrivilege": false,
        "source": "nvd@nist.gov",
        "type": "Primary",
        "userInteractionRequired": false
      }
    ],
    "cvssMetricV30": [
      {
        "cvssData": {
          "attackComplexity": "LOW",
          "attackVector": "NETWORK",
          "availabilityImpact": "NONE",
          "baseScore": 7.5,
          "baseSeverity": "HIGH",
          "confidentialityImpact": "HIGH",
          "integrityImpact": "NONE",
          "privilegesRequired": "NONE",
          "scope": "UNCHANGED",
          "userInteraction": "NONE",
          "vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
          "version": "3.0"
        },
        "exploitabilityScore": 3.9,
        "impactScore": 3.6,
        "source": "nvd@nist.gov",
        "type": "Primary"
      }
    ]
  },
  "published": "2019-03-25T16:29:00.303",
  "references": [
    {
      "source": "ics-cert@hq.dhs.gov",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
    },
    {
      "source": "af854a3a-2127-422b-91ae-364da2661108",
      "tags": [
        "Mitigation",
        "Third Party Advisory",
        "US Government Resource"
      ],
      "url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01"
    }
  ],
  "sourceIdentifier": "ics-cert@hq.dhs.gov",
  "vulnStatus": "Modified",
  "weaknesses": [
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-312"
        }
      ],
      "source": "ics-cert@hq.dhs.gov",
      "type": "Secondary"
    },
    {
      "description": [
        {
          "lang": "en",
          "value": "CWE-200"
        }
      ],
      "source": "nvd@nist.gov",
      "type": "Primary"
    }
  ]
}



Log in or create an account to share your comment.




Tags
Taxonomy of the tags.


Loading…

Loading…

Loading…

Forecast uses a logistic model when the trend is rising, or an exponential decay model when the trend is falling. Fitted via linearized least squares.

Sightings

Author Source Type Date Other

Nomenclature

  • Seen: The vulnerability was mentioned, discussed, or observed by the user.
  • Confirmed: The vulnerability has been validated from an analyst's perspective.
  • Published Proof of Concept: A public proof of concept is available for this vulnerability.
  • Exploited: The vulnerability was observed as exploited by the user who reported the sighting.
  • Patched: The vulnerability was observed as successfully patched by the user who reported the sighting.
  • Not exploited: The vulnerability was not observed as exploited by the user who reported the sighting.
  • Not confirmed: The user expressed doubt about the validity of the vulnerability.
  • Not patched: The vulnerability was not observed as successfully patched by the user who reported the sighting.

Loading…

Detection rules are retrieved from Rulezet.

Loading…

Loading…

Related by attack behaviour

Vulnerabilities whose description is nearest to this one in the vector space of the CIRCL/vulnerability-attack-technique-biencoder model. This is a similarity search over the bi-encoder space (plain cosine), not a classification, and it has no measured accuracy.


Loading…